Performs external calibration adjustment on a DSA device with the specified gain and coupling configuration.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

reference voltage

reference voltage specifies in volts the known voltage to use as a reference for calibration. This unit is Vrms for AC coupling or VDC for DC coupling.

datatype_icon

gain

gain specifies the gain setting to calibrate, in decibels.

datatype_icon

coupling

coupling specifies the coupling setting to calibrate.

Name Description
AC

AC coupling.

DC

DC coupling.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.