FieldDAQ Calibration

Version:
    Last Modified: April 18, 2018

    The FD-11613 and FD-11614 devices use software calibration to adjust the software scaling of signals read from and produced by your device. Using a precise reference signal, your device measures and calculates scaling constants for analog input. The scaling constants are stored in nonvolatile memory (EEPROM) on your device. Frequent calibration produces the most stable and repeatable measurement performance. The device is not harmed in any way if you recalibrate it often.

    spd-note-note
    Note  
    • Calibrating your device takes some time. Do not be alarmed if the Adjust FieldDAQ Calibration function/VI takes several seconds to execute.
    • For best results, stop any ongoing tasks and disconnect any unnecessary external connections before running calibration.

    The FD-11613 and FD-11614 devices do not self-calibrate or automatically calibrate. You must use a manual procedure to calibrate these devices.