    Specifies whether to apply a digital filter to the digital output of the analog triggering circuitry (the Analog Comparison Event). When enabled, the analog signal must stay within the trigger window for the minimum pulse width before being recognized. Use filtering for noisy trigger signals that transition in and out of the window rapidly.