From 6:00 PM CST Friday, Feb 15th - 2:00 AM CST Sunday, Feb 17th, ni.com will be undergoing system upgrades that may result in temporary service interruption.

We appreciate your patience as we improve our online experience.

Table Of Contents

Platinum RTD Types

Last Modified: November 16, 2017

The following table lists common platinum RTD types and standards. All of these RTD types are supported in NI-DAQmx. Notice that there are some shared standards. The TCR and the Callendar-Van Dusen coefficients are more important than the standards.

Standards Material TCR Typical R0 (Ω) Callendar-Van Dusen Coefficient Notes
  • IEC-751
  • DIN 43760
  • BS 1904
  • ASTM-E1137
  • EN-60751
  • IEC-60751
Platinum 3851
  • 100 Ω
  • 1000 Ω
  • A = 3.9083 × 10-3
  • B = -5.775 × 10-7
  • C = -4.183 × 10-12
Most common RTDs
Low-cost vendor compliant RTD[1] Platinum 3750 1000 Ω
  • A = 3.81 × 10-3
  • B = -6.02 × 10-7
  • C = -6.0 × 10-12
Low-cost RTD
JISC 1604 Platinum 3916 100 Ω
  • A = 3.9739 × 10-3
  • B = -5.870 × 10-7
  • C = -4.4 × 10-12
Used primarily in Japan
US Industrial Standard D-100 American Platinum 3920 100 Ω
  • A = 3.9787 × 10-3
  • B = -5.8686 × 10-7
  • C = -4.167 × 10-12
Low-cost RTD
US Industrial Standard American Platinum 3911 100 Ω
  • A = 3.9692 × 10-3
  • B = -5.8495 × 10-7
  • C = -4.233 × 10-12
Low-cost RTD
ITS-90 Platinum 3928 100 Ω
  • A = 3.9888 × 10-3
  • B = -5.915 × 10-7
  • C = -3.85 × 10-12
The definition of temperature
  • 1 No standard. Check the TCR.

Recently Viewed Topics