Table Of Contents

Platinum RTD Types

Version:
    Last Modified: June 22, 2016

    The following table lists common platinum RTD types and standards. All of these RTD types are supported in NI-DAQmx. Notice that there are some shared standards. The TCR and the Callendar-Van Dusen coefficients are more important than the standards.

    Standards Material TCR Typical R0 (Ω) Callendar-Van Dusen Coefficient Notes
    • IEC-751
    • DIN 43760
    • BS 1904
    • ASTM-E1137
    • EN-60751
    • IEC-60751
    Platinum 3851
    • 100 Ω
    • 1000 Ω
    • A = 3.9083 × 10-3
    • B = -5.775 × 10-7
    • C = -4.183 × 10-12
    Most common RTDs
    Low-cost vendor compliant RTD[1] Platinum 3750 1000 Ω
    • A = 3.81 × 10-3
    • B = -6.02 × 10-7
    • C = -6.0 × 10-12
    Low-cost RTD
    JISC 1604 Platinum 3916 100 Ω
    • A = 3.9739 × 10-3
    • B = -5.870 × 10-7
    • C = -4.4 × 10-12
    Used primarily in Japan
    US Industrial Standard D-100 American Platinum 3920 100 Ω
    • A = 3.9787 × 10-3
    • B = -5.8686 × 10-7
    • C = -4.167 × 10-12
    Low-cost RTD
    US Industrial Standard American Platinum 3911 100 Ω
    • A = 3.9692 × 10-3
    • B = -5.8495 × 10-7
    • C = -4.233 × 10-12
    Low-cost RTD
    ITS-90 Platinum 3928 100 Ω
    • A = 3.9888 × 10-3
    • B = -5.915 × 10-7
    • C = -3.85 × 10-12
    The definition of temperature
    • 1 No standard. Check the TCR.