Half-Bridge Type I

Version:
    Last Modified: February 16, 2016

    The following figure shows how to position strain gage resistors in an axial configuration for the half-bridge type I.

    The following figure shows how to position strain gage resistors in a bending configuration for the half-bridge type I.

    Half-bridge type I strain gage configurations have the following characteristics:
    • Two active strain gage elements, one mounted in the direction of axial strain and the other acting as a Poisson gage and mounted transverse, or perpendicular, to the principal axis of strain.
    • Completion resistors which provide half-bridge completion.
    • Sensitivity to both axial and bending strain.
    • Compensation for temperature.
    • Compensation for the aggregate effect on the principle strain measurement due to the Poisson's ratio of the material.
    • Sensitivity at 1000 µε is ~ 0.65 mVout/ VEX input

    Half-Bridge Type I Circuit Diagram

    The following symbols apply to the circuit diagram:
    • R1 is the half-bridge completion resistor.
    • R2 is the half-bridge completion resistor.
    • R3 is the active strain gage element measuring compression due to the Poisson effect (–ε).
    • R4 is the active strain gage element measuring tensile strain (+ε).
    • VEX is the excitation voltage.
    • RL is the lead resistance.
    • VCH is the measured voltage.

    The following equation converts voltage ratios to strain units for half-bridge type I configurations.

    where Vr is the voltage ratio that virtual channels use in the voltage-to-strain conversion equation, GF is the gage factor, v is the Poisson's ratio, RL is the lead resistance, and Rg is the nominal gage resistance.