DAQmx Perform Thermocouple Lead Offset Nulling Calibration (G Dataflow)

Version:
    Last Modified: February 6, 2017

    Performs thermocouple lead offset nulling calibration, on the channels in the task, to compensate for offsets introduced by open thermocouple detection.

    Keep the measured temperature as constant as possible while performing this adjustment.

    connector_pane_image
    datatype_icon

    task in

    task in is the name of the task that the operation applies.

    datatype_icon

    channels

    channels is a subset of virtual channels in the task that you want to calibrate. Use this input if you do not want to calibrate all the channels in the task or if some channels in the task have open thermocouple detection disabled. If the input is empty, this node attempts to calibrate all virtual channels in the task.

    datatype_icon

    error in

    error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

    datatype_icon

    skip unsupported channels

    skip unsupported channels specifies whether or not to skip channels that do not support calibration. If skip unsupported channels is TRUE, this node calibrates only supported channels. If FALSE, this node calibrates the channels specified by channels. The default is FALSE.

    datatype_icon

    error out

    error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.

    datatype_icon

    task out

    task out is a reference to the task after this node runs.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: LabVIEW NXG does not support FPGA devices