Table Of Contents

DAQmx Setup SCXI Calibration (1124) (G Dataflow)

Last Modified: February 6, 2017

Writes the specified binary value to the DAC on the specified physical channel at the specified range.

spd-note-note
Note  

Specify at least two calibration points for each physical channel and range you calibrate. Use 0 and 4095 for voltage calibration ranges. Use 255 and 4095 for current calibration ranges.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

physical channel

physical channel specifies the physical channel to calibrate.

datatype_icon

range

range specifies the range to calibrate.

Name Description
0Volts to 1Volts

0 V to 1 V.

0Volts to 5Volts

0 V to 5 V.

0Volts to 10Volts

0 V to 10 V.

-1Volts to 1Volts

-1 V to 1 V.

-5Volts to 5Volts

-5 V to 5 V.

-10Volts to 10Volts

-10 V to 10 V.

0Amps to 20mAmps

0 A to 20 mA.

datatype_icon

DAC value

DAC value is the binary number to write to the DAC.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.