Table Of Contents

DAQmx Setup C Series Calibration (9218) (G Dataflow)

Last Modified: February 6, 2017

Sets up external calibration for an NI 9218 device.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

physical channels

physical channels specifies the name of the physical channel to use to create the virtual channel. The DAQmx physical channel constant lists all physical channels on devices and modules installed in the system.

You also can wire a string that contains a physical channel to this input. You cannot specify more than one physical channel.

datatype_icon

measType

measType specifies the type of measurement for the device.

Name Description
Voltage

Voltage.

Current

Current.

Bridge

Wheatstone bridge.

datatype_icon

rangeMax

rangeMax specifies the maximum value in the range, in the units of the measurement type specified by measType. The units for the range inputs should correspond to the measurement type.

datatype_icon

rangeMin

rangeMin specifies the minimum value in the range, in the units of the measurement type specified by measType. The units for the range inputs should correspond to the measurement type.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.

Where This Node Can Run:

Desktop OS: Windows

FPGA: LabVIEW NXG does not support FPGA devices