DAQmx Self Calibrate (G Dataflow)

Version:
    Last Modified: February 6, 2017

    Measures the onboard reference voltage of the device, and adjusts the self-calibration constants, to account for any errors caused by short-term fluctuations in the operating environment. When you self-calibrate a device, no external signal connections are necessary.

    For 42xx devices, this node performs a self-calibration on the 4200 baseboard, then performs an offset-only calibration on the full signal path of the device.

    connector_pane_image
    datatype_icon

    dev in

    dev in is the name as configured in MAX of the device to which this operation applies. A DAQmx device name constant lists all devices installed in the system.

    datatype_icon

    error in

    error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

    datatype_icon

    dev out

    dev out is the name of the device the operation applied to.

    datatype_icon

    error out

    error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.