Table Of Contents

DAQmx Adjust DSA Calibration (Analog Output » Analog Output) (G Dataflow)

Version:
    Last Modified: February 6, 2017

    Adjusts the external calibration constants for the analog output section of a DSA device.

    You must use the device to generate a high voltage and low voltage at a specified gain, measure the high and low voltages, then specify the requested high voltage, low voltage, and gain along with the actual high voltage and low voltage.

    connector_pane_image
    datatype_icon

    calHandle in

    calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

    datatype_icon

    channel

    channel is the number of the channel to calibrate. This number is the numeric portion of the physical channel name, not the full physical channel name.

    datatype_icon

    requested low voltage

    requested low voltage is the low voltage you attempted to generate at the gain setting you specified.

    datatype_icon

    actual low voltage

    actual low voltage is the actual low voltage an external sensor measures.

    datatype_icon

    error in

    error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

    datatype_icon

    requested high voltage

    requested high voltage is the high voltage you attempted to generate at the gain setting you specified.

    datatype_icon

    actual high voltage

    actual high voltage is the actual high voltage an external sensor measures.

    datatype_icon

    gain setting

    gain setting is the gain setting you used when you attempted to generate the requested high voltage and requested low voltage.

    datatype_icon

    calHandle out

    calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

    datatype_icon

    error out

    error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: LabVIEW NXG does not support FPGA devices


    Recently Viewed Topics