Table Of Contents

DAQmx Adjust SC Express Calibration (4330/1) (G Dataflow)

Last Modified: February 6, 2017

Adjusts the external calibration constants for an NI PXIe-4330 or NI PXIe-4331 device.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

reference voltage

reference voltage is the voltage across the AI+ and AI- terminals, measured using a DMM.

datatype_icon

reference excitation

reference excitation is the voltage across the EX+ and EX- terminals, measured using a second DMM.

datatype_icon

shunt location

shunt location specifies the resistive leg of the Wheatstone bridge that has an external shunt resistor connected in parallel.

Name Description
R3

R3.

R4

R4.

None

No shunt resistor connected.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.

Where This Node Can Run:

Desktop OS: Windows

FPGA: LabVIEW NXG does not support FPGA devices


Recently Viewed Topics