The NI 5772 allows you to configure the clock, ADC, sampling, and other settings with the User Command, User Data 0, and User Data 1 signals.

User Setting User Command User Data 0 Value User Data 1 Value User Return Value
Clock Selection Enumeration 0 0 = Internal Sample clock

1 = Internal Sample clock locked to an external Reference clock through the CLK IN connector

2 = External Sample clock through the CLK IN connector (disables the CLK OUT connector)

3 = Internal Sample clock locked to an external Reference clock through Sync Clock

N/A N/A
ADC Self-Calibration 1 Performs a self-calibration on the ADC12D800RF. N/A N/A
Set TIS Mode 4 0 = TIS mode disabled

1 = TIS mode (Channel 0 only) enabled

2 = TIS mode (Channel 1 only) enabled

3 = Dual-TIS Mode (Channels 0 and 1) enabled

Note  

DC-coupled modules do not support Dual-TIS Mode.

N/A N/A
Count Metastability 5 Issues 1,000 timestamp measurements and returns the number of timestamps that are metastable. N/A N/A
Set Timestamp Source 6 Selects the source of the timestamp measurement. N/A N/A
Read Revision 240 Reads the CLIP revision number. N/A N/A
Read Oldest Compatible Revision 242 Reads the CLIP Oldest Compatible Revision number. N/A N/A
Reinitialize 255 Configures the NI 5772 to its default power-on configuration. N/A N/A