Configures the properties that control the function generator's hold trigger. These properties are the hold trigger source and slope.
A hold trigger pauses generation. From the paused state, a resume trigger resumes generation, a stop trigger terminates generation, and start trigger behavior is vendor defined.
Specifies the slope of the hold trigger. The driver sets the IviFgen Hold Trigger Slope [HT] property to this value. Valid Values: IVIFGEN_VAL_TRIGGER_POSITIVE (0) - Triggers on a positive slope. IVIFGEN_VAL_TRIGGER_NEGATIVE (1) - Triggers on a negative slope. IVIFGEN_VAL_TRIGGER_EITHER (2) - Triggers on either a positive or negative slope. Default: IVIFGEN_VAL_TRIGGER_POSITIVE (0) - Triggers on a positive slope.
The instrument handle that you obtain from the IviFgen Initialize or IviFgen Initialize With Options Node. The handle identifies a particular instrument session.
Default: None
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Default: No error
Specifies the virtual channel name that you assign to the instrument in the Configuration Utility. Users who want to achieve interchangeability should use a virtual channel name. The virtual channel name should be sufficiently specific to the test system such that it is unlikely to conflict with a physical channel name. Virtual channel names are aliases for instrument-specific channel strings. The instrument-specific channel strings can differ from one instrument to another. Virtual channel names allow you to use and swap instruments without having to change the channel names in your source code. You assign a virtual channel name to an instrument-specific channel through the Configuration Utility. This control accepts virtual channel names you have assigned to the specific instrument you are using. It also accepts the instrument-specific channel names.
Default: "CHANNEL1"
The instrument handle that you obtain from the IviFgen Initialize or IviFgen Initialize With Options Node. The handle identifies a particular instrument session.
Default: None
Specifies the source of the hold trigger. The driver sets the IviFgen Hold Trigger Source [HT] property to this value. Valid Values: IVIFGEN_VAL_TRIGGER_SOURCE_NONE ("") - No trigger source IVIFGEN_VAL_TRIGGER_SOURCE_IMMEDIATE ("Immediate") - Trigger immediately IVIFGEN_VAL_TRIGGER_SOURCE_EXTERNAL ("External") - External trigger source IVIFGEN_VAL_TRIGGER_SOURCE_INTERNAL ("Internal") - Internal trigger source IVIFGEN_VAL_TRIGGER_SOURCE_SOFTWARE ("Software") - Software trigger IVIFGEN_VAL_TRIGGER_SOURCE_LAN0 ("LAN0") - LAN0 IVIFGEN_VAL_TRIGGER_SOURCE_LAN1 ("LAN1") - LAN1 IVIFGEN_VAL_TRIGGER_SOURCE_LAN2 ("LAN2") - LAN2 IVIFGEN_VAL_TRIGGER_SOURCE_LAN3 ("LAN3") - LAN3 IVIFGEN_VAL_TRIGGER_SOURCE_LAN4 ("LAN4") - LAN4 IVIFGEN_VAL_TRIGGER_SOURCE_LAN5 ("LAN5") - LAN5 IVIFGEN_VAL_TRIGGER_SOURCE_LAN6 ("LAN6") - LAN6 IVIFGEN_VAL_TRIGGER_SOURCE_LAN7 ("LAN7") - LAN7 IVIFGEN_VAL_TRIGGER_SOURCE_LXI0 ("LXI0") - LXI Trigger Bus Line 0 IVIFGEN_VAL_TRIGGER_SOURCE_LXI1 ("LXI1") - LXI Trigger Bus Line 1 IVIFGEN_VAL_TRIGGER_SOURCE_LXI2 ("LXI2") - LXI Trigger Bus Line 2 IVIFGEN_VAL_TRIGGER_SOURCE_LXI3 ("LXI3") - LXI Trigger Bus Line 3 IVIFGEN_VAL_TRIGGER_SOURCE_LXI4 ("LXI4") - LXI Trigger Bus Line 4 IVIFGEN_VAL_TRIGGER_SOURCE_LXI5 ("LXI5") - LXI Trigger Bus Line 5 IVIFGEN_VAL_TRIGGER_SOURCE_LXI6 ("LXI6") - LXI Trigger Bus Line 6 IVIFGEN_VAL_TRIGGER_SOURCE_LXI7 ("LXI7") - LXI Trigger Bus Line 7 IVIFGEN_VAL_TRIGGER_SOURCE_TTL0 ("TTL0") - TTL Interface 0 IVIFGEN_VAL_TRIGGER_SOURCE_TTL1 ("TTL1") - TTL Interface 1 IVIFGEN_VAL_TRIGGER_SOURCE_TTL2 ("TTL2") - TTL Interface 2 IVIFGEN_VAL_TRIGGER_SOURCE_TTL3 ("TTL3") - TTL Interface 3 IVIFGEN_VAL_TRIGGER_SOURCE_TTL4 ("TTL4") - TTL Interface 4 IVIFGEN_VAL_TRIGGER_SOURCE_TTL5 ("TTL5") - TTL Interface 5 IVIFGEN_VAL_TRIGGER_SOURCE_TTL6 ("TTL6") - TTL Interface 6 IVIFGEN_VAL_TRIGGER_SOURCE_TTL7 ("TTL7") - TTL Interface 7 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_STAR ("PXI_STAR") - PXI Star Interface IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG0 ("PXI_TRIG0") - PXI Trigger Bus Line 0 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG1 ("PXI_TRIG1") - PXI Trigger Bus Line 1 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG2 ("PXI_TRIG2") - PXI Trigger Bus Line 2 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG3 ("PXI_TRIG3") - PXI Trigger Bus Line 3 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG4 ("PXI_TRIG4") - PXI Trigger Bus Line 4 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG5 ("PXI_TRIG5") - PXI Trigger Bus Line 5 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG6 ("PXI_TRIG6") - PXI Trigger Bus Line 6 IVIFGEN_VAL_TRIGGER_SOURCE_PXI_TRIG7 ("PXI_TRIG7") - PXI Trigger Bus Line 7 IVIFGEN_VAL_TRIGGER_SOURCE_PXIE_DSTARA ("PXIe_DSTARA") - PXI Express DStar Line A IVIFGEN_VAL_TRIGGER_SOURCE_PXIE_DSTARB ("PXIe_DSTARB") - PXI Express DStar Line B IVIFGEN_VAL_TRIGGER_SOURCE_PXIE_DSTARC ("PXIe_DSTARC") - PXI Express DStar Line C IVIFGEN_VAL_TRIGGER_SOURCE_RTSI0 ("RTSI0") - RTSI Bus Line 0 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI1 ("RTSI1") - RTSI Bus Line 1 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI2 ("RTSI2") - RTSI Bus Line 2 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI3 ("RTSI3") - RTSI Bus Line 3 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI4 ("RTSI4") - RTSI Bus Line 4 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI5 ("RTSI5") - RTSI Bus Line 5 IVIFGEN_VAL_TRIGGER_SOURCE_RTSI6 ("RTSI6") - RTSI Bus Line 6 Default: IVIFGEN_VAL_TRIGGER_SOURCE_NONE ("") - No trigger source
Error information.
The node produces this output according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.