IviDmm Configure Fixed Ref Junction [TC] (G Dataflow)

Version:
    Last Modified: October 24, 2017

    Configures the fixed reference junction for a thermocouple with a fixed reference junction type.

    This node affects the behavior of the instrument only when the IviDmm>Temperature Measurements>Thermocouple>Reference Junction Type [TC] property is set to IVIDMM_VAL_TEMP_REF_JUNC_FIXED.

    connector_pane_image
    datatype_icon

    fixed ref junc

    Pass the reference junction temperature when a fixed reference junction type thermocouple is used to take the temperature measurement. Units: Degrees Celsius. The driver sets the IviDmm>Temperature Measurements>Thermocouple>Fixed Reference Junction [TC] property to this value.

    spd-note-note
    Note  

    This property may also be used to specify the thermocouple junction temperature of an instrument that does not have an internal temperature sensor.

    Default: 0.0

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.
    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    instrument handle out

    The instrument handle that you obtain from the IviDmm Initialize Node. The handle identifies a particular instrument session.

    Default: None

    datatype_icon

    instrument handle

    The instrument handle that you obtain from the IviDmm Initialize Node. The handle identifies a particular instrument session.

    Default: None

    Extension Group

    This node is part of the IviDmmThermocouple [TC] extension group.