Table Of Contents

IviDigitizer Configure Glitch Trigger Source [GT] (G Dataflow)

Version:
    Last Modified: October 24, 2017

    Configures the glitch trigger. A glitch trigger occurs when the trigger signal has a pulse with a width that is less than or greater than the glitch width.

    The end user specifies which comparison criterion to use with the Glitch Arm Condition input. The end-user specifies the glitch width with the Glitch Arm Width input. The end-user specifies the polarity of the pulse with the Glitch Arm Polarity input.

    The trigger does not actually occur until the edge of a pulse that corresponds to the Glitch Arm Width input and Glitch Arm Polarity input crosses the threshold the end-user specifies in the Trigger Level input. This node affects instrument behavior only if the IviDigitizer Trigger Type property is set to glitch trigger.

    spd-note-note
    Note  

    Set the trigger type and trigger coupling before calling this node.

    spd-note-note
    Note  

    This node is part of the IviDigitizerGlitchTrigger [GT] extension group.

    connector_pane_image
    datatype_icon

    Glitch Polarity (Positive)

    Specifies the glitch polarity. This value sets the IviDigitizer Glitch Arm Polarity property. Valid Values: - IVIDIGITIZER_VAL_GLITCH_POSITIVE (1) The digitizer triggers on a positive glitch. - IVIDIGITIZER_VAL_GLITCH_NEGATIVE (2) The digitizer triggers on a negative glitch. - IVIDIGITIZER_VAL_GLITCH_EITHER (3) The digitizer triggers on either a positive or negative glitch. Default Value: IVIDIGITIZER_VAL_GLITCH_POSITIVE (1)

    datatype_icon

    Trigger Level (1.0)

    Specifies the trigger level. This value sets the IviDigitizer Trigger Level property. Unit: volt Default Value: 1.0

    datatype_icon

    Trigger Source ("")

    Specifies the trigger source. This value sets the trigger source property. Users who want to achieve interchangeability should use a virtual trigger source name. The virtual trigger source name should be sufficiently specific to the test system such that it is unlikely to conflict with a physical trigger source name. Default Value: "" Note: You can specify the trace name as a string variable or as a literal enclosed in double quotes.

    datatype_icon

    Glitch Width (0.1)

    Specifies the glitch arming glitch width, in seconds. This value sets the IviDigitizer Glitch Arm Width property. Unit: second Default Value: 0.1

    datatype_icon

    Glitch Condition (Less Than)

    Specifies the glitch condition. This value sets the IviDigitizer Glitch Arm Condition property. Valid Values: - IVIDIGITIZER_VAL_GLITCH_LESS_THAN (1) The digitizer triggers when the pulse width is less than the value you specify with the IviDigitizer Glitch Arm Width property. - IVIDIGITIZER_VAL_GLITCH_GREATER_THAN (2) The digitizer triggers when the pulse width is greater than the value you specify with the IviDigitizer Glitch Arm Width property. Default Value: IVIDIGITIZER_VAL_GLITCH_LESS_THAN (1)

    datatype_icon

    instrument handle out

    The instrument handle that you obtain from the IviDigitizer Initialize Node or IviDigitizer Initialize With Options Node. The handle identifies a particular instrument session.

    Default: None

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.
    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    instrument handle

    The instrument handle that you obtain from the IviDigitizer Initialize Node or IviDigitizer Initialize With Options Node. The handle identifies a particular instrument session.

    Default: None

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported

    Web Server: Not supported in VIs that run in a web application