    The NI 5771 allows you to configure the clock, ADC, sampling, and other settings with the User Command, User Data 0, and User Data 1 signals.

    User Setting User Command User Data 0 Value User Data 1 Value User Return Value
    Clock Settings 0 0 = Internal Sample clock

    1 = Internal Sample clock locked to an external Reference clock through the CLK IN connector

    2 = External Sample clock through the CLK IN connector (disables the CLK OUT connector)

    3 = Internal Sample clock locked to an external Reference clock through Sync Clock

    N/A N/A
    ADC Self-Calibration 1 N/A N/A N/A
    ADC Full Scale Range 2 Gain Setting = 0x0000 to 0x01FF

    0x0000 = 560 mV, Minimum

    0x0100 = 700 mV, Nominal

    0x01FF = 840 mV, Maximum

    0 = AI 0

    1 = AI 1

    N/A
    ADC Offset 3 Positive Offset = 0x0000 to 0x00FF

    Negative Offset = 0x0100 to 0x01FF

    (0.176 mV per LSB)

    0 = AI 0

    1 = AI 1

    N/A
    TIS Mode 4 0 = TIS mode disabled

    1 = TIS mode enabled

    0 = AI 0

    1 = AI 1

    N/A
    Frequency Adjust DAC Value 5 DAC value = 0x0000 to 0xFFFF N/A N/A
    Read User EEPROM 6 EEPROM address to read data. N/A Read data
    Write User EEPROM 7 EEPROM address to write data. Write data N/A
    Exported Sample Clock 8 0 = CLK OUT disabled

    1 = CLK OUT enabled

    Note: If you set this signal to a nonzero value and your device is already configured to use the external Sample clock through the CLK IN connector, the User Command Status signal returns a nonzero value indicating that the command was unsuccessful.

    N/A N/A
    Reinitialize 0xFF N/A N/A N/A